NASA Logo

NTRS

NTRS - NASA Technical Reports Server

Back to Results
Selection and use of TLDS for high precision NERVA shielding measurementsAn experimental evaluation of thermoluminescent dosimeters was performed in order to select high precision dosimeters for a study whose purpose is to measure gamma streaming through the coolant passages of a simulated flight type internal NERVA reactor shield. Based on this study, the CaF2 chip TLDs are the most reproducible dosimeters with reproducibility generally within a few percent, but none of the TLDs tested met the reproducibility criterion of plus or minus 2%.
Document ID
19720010010
Acquisition Source
Legacy CDMS
Document Type
Conference Paper
Authors
Woodsum, H. C.
(Westinghouse Electric Corp. Pittsburgh, PA, United States)
Date Acquired
August 6, 2013
Publication Date
January 1, 1972
Publication Information
Publication: NASA, Washington Proc. of the Natl. Symp. on Nat. and Manmade Radiation in Space p466-470 (SEE N72-17601 08-22)
Subject Category
Nuclear Engineering
Accession Number
72N17660
Distribution Limits
Public
Copyright
Work of the US Gov. Public Use Permitted.