A CCD scanned (Hg, Cd)Te array for earth viewing applicationsA 9 channel demonstration of 10 micrometer photoconductive mercury cadmium telluride detector array, bipolar preamplifiers, and a Si CCD multiplexer, integrated into a focal plane assembly and operating at 105 K and at 17 Hz is reported. The result of a direct coupling experiment between a 5-micrometer photovoltaic mercury cadmium telluride photodiode and a CCD is revealed. No appreciable degradation in the signal-to-noise ratio was experienced.
Document ID
19770010347
Acquisition Source
Legacy CDMS
Document Type
Conference Paper
Authors
Iwasa, S.
(Honeywell Radiation Center Lexington, MA, United States)
White, W. J.
(Honeywell Radiation Center Lexington, MA, United States)
Date Acquired
August 8, 2013
Publication Date
January 1, 1976
Publication Information
Publication: JPL Conf. on Charge-Coupled Device Technol. and Appls.
Subject Category
Instrumentation And Photography
Accession Number
77N17290
Funding Number(s)
CONTRACT_GRANT: NAS5-22339
Distribution Limits
Public
Copyright
Work of the US Gov. Public Use Permitted.