NASA Logo

NTRS

NTRS - NASA Technical Reports Server

Back to Results
The CCN counter of the IOPG of Puy de Dome: Main characterisics and results of measurementsIn performing the data analysis of the measurements achieved during the Workshop with the device, a global approach was preferred rather than an individual analysis, in order to illustrate some main characteristics in the behavior of the device with respect to a mean behavior resulting from a general survey of all the equipments involved in each experiment. The device tends generally to overestimate the CCN concentrations measured near the high supersaturations and sometimes underestimates the concentrations close to 0.1% or 0.2% of supersaturation. Despite the fact that it belongs to a type of static diffusion chamber, it shows, however, similar spectra to those obtained with other types of chambers (continuous flow diffusion chamber and haze chamber).
Document ID
19820007513
Acquisition Source
Legacy CDMS
Document Type
Conference Paper
Authors
Serpolay, R.
(Clermont-Ferrand Univ.)
Date Acquired
August 10, 2013
Publication Date
November 1, 1981
Publication Information
Publication: Desert Research Inst. The Third Intern. Cloud Condensation Nuclei Workshop
Subject Category
Instrumentation And Photography
Accession Number
82N15386
Distribution Limits
Public
Copyright
Work of the US Gov. Public Use Permitted.