A sealed titanium window proportional counter for the detection of .5-keV X rays.A sealed Ti window proportional counter sensitive to X radiation in the energy range 0.35-0.45 keV and above 1.5 keV is described. Measurements of the Ti mass absorption coefficients and a graphical summary of the literature values are presented. For a proportional counter with a 930 microgram/sq cm (2.1-micron) Ti window, the peak efficiency at 0.45 keV is found to lie between 4.6% and 7.1%. An application in X-ray astronomy involving a rocket observation of Sco X-1 is discussed.