Cleaning of contaminated channel electron multiplier arrays.Polymerized organic materials found in vacuum systems frequently contaminate the individual channels (which can be as small as 15 microns) of an oil-pumped channel electron multiplier array used as a vacuum UV- to visible-radiation conversion device. It is shown that exposure to an oxygen plasma is an effective means of decontaminating the channels, thereby restoring the sensitivity of the device.