NASA Logo

NTRS

NTRS - NASA Technical Reports Server

Back to Results
Active cavity radiometer.The active cavity radiometer (ACR) is a pyrheliometer that accurately defines the absolute radiation scale. The physics of the pyrheliometric method and the ACR approach to this method are presented in detail. A mathematical abstraction of the method is generated through a quasi-equilibrium analysis of the power balance of the ACR's cavity detector. An error analysis is carried out on the quasi-equilibrium equation to determine the uncertainties of ACR measurements relative to the absolute radiation scale. The uncertainty of ACR measurements as a function of irradiance level is presented in graphical form.
Document ID
19730041434
Acquisition Source
Legacy CDMS
Document Type
Reprint (Version printed in journal)
Authors
Willson, R. C.
(California Institute of Technology, Jet Propulsion Laboratory, Pasadena Calif., United States)
Date Acquired
August 7, 2013
Publication Date
April 1, 1973
Publication Information
Publication: Applied Optics
Volume: 12
Subject Category
Instrumentation And Photography
Accession Number
73A26236
Funding Number(s)
CONTRACT_GRANT: NAS7-100
Distribution Limits
Public
Copyright
Other

Available Downloads

There are no available downloads for this record.