Use of channel electron multipliers as secondary standard detectors at EUV wavelengthsThe procedures available for photometric calibration at extreme ultraviolet (EUV) wavelengths are outlined and the requirements for a secondary standard EUV photomultiplier defined. The performance of a number of commercially available channel electron multipliers over the 304-1350-A wavelength range is described, and their suitability for use as secondary standards is discussed in detail. Although none of the multipliers evaluated fully met the requirements for a secondary standard, it proved possible to calibrate absolutely a Mullard cone channel over the required wavelength range to an accuracy of plus or minus 9% and to employ it as a secondary standard in the calibration of a series of sounding rocket spectrometers.