NASA Logo

NTRS

NTRS - NASA Technical Reports Server

Back to Results
Quantum efficiency of a channel electron multiplier in the far ultravioletVariation of the quantum efficiency of a channel electron multiplier (CEM) in the wavelength range from 1200 to 2536 A is studied. Emphasis is on measurement of CEM sensitivity longward of 1500 A. Results indicate an overall rapid decrease in quantum efficiency with increasing wavelength, with little evidence for a possible change in slope in the range from 2000 to 2500 A. The lowest efficiency measured is 4.5 + or -2.5 times 10 to the minus ninth count/photon at 2536 A. These efficiencies should ensure that unwanted radiation longward of 1500 A can be effectively removed from the bandpass of the instrument.
Document ID
19760032885
Acquisition Source
Legacy CDMS
Document Type
Reprint (Version printed in journal)
Authors
Paresce, F.
(California, University Berkeley, Calif., United States)
Date Acquired
August 8, 2013
Publication Date
December 1, 1975
Publication Information
Publication: Applied Optics
Volume: 14
Subject Category
Instrumentation And Photography
Accession Number
76A15851
Funding Number(s)
CONTRACT_GRANT: NAS9-13799
Distribution Limits
Public
Copyright
Other

Available Downloads

There are no available downloads for this record.
No Preview Available