NASA Logo

NTRS

NTRS - NASA Technical Reports Server

Back to Results
An automated secondary standard for calibrating liquid flowmetersA secondary working standard of flow calibration has been developed to be used in place of a primary weight-time standard, and which can thereby effect a 75 percent reduction in calibration time while maintaining acceptable accuracies. The secondary standard uses six previously calibrated turbine-type flowmeters built into two manifold systems containing automatically switched flow valves. The pair of systems is capable of covering the flow range of 0.0004 to 19 1/s (0.007 to 300 gpm) with the uncertainty in volume flow rate not exceeding plus or minus 0.25 percent over the range 0.06 to 19 1/s and not exceeding plus or minus 0.5 percent over the range 0.0004 to 0.06 1/s. Data reduction and plotting of results are by computer.
Document ID
19770043945
Acquisition Source
Legacy CDMS
Document Type
Conference Proceedings
Authors
Hobart, H. F.
(NASA Lewis Research Center Cleveland, Ohio, United States)
Date Acquired
August 8, 2013
Publication Date
January 1, 1976
Subject Category
Instrumentation And Photography
Meeting Information
Meeting: International Instrumentation Symposium on Advances in test measurement
Location: San Diego, CA
Start Date: May 25, 1976
End Date: May 27, 1976
Accession Number
77A26797
Distribution Limits
Public
Copyright
Other

Available Downloads

There are no available downloads for this record.
No Preview Available