Which calibration-pulse location method is robust?The Threshold method, with high threshold and contiguous-block parameter values, was found to be a robust method of locating calibration pulses in the presence of light-leak and shutter-edge pulses within the calibration window. Tests used digitized calibration-window background and light-pulse data from the Landsat-5 Thematic Mapper (TM) instrument, analyzed by special software on an Apple II+ personal computer and on a VAX 11/780 minicomputer.
Document ID
19860052267
Acquisition Source
Legacy CDMS
Document Type
Conference Paper
Authors
Gunther, F. J. (Computer Sciences Corp. Silver Spring, MD, United States)