An ac bridge readout for bolometric detectorsThe authors have developed a bolometer readout circuit which greatly improves the low-frequency stability of bolometric detectors. The circuit uses an ac bias voltage and two matched bolometers and allows stable dc bolometer operation for integration times greater than 10 s. In astronomical applications the readout allows for qualitatively different observation modes (e.g. staring or slow-drift scanning) which are particularly well suited for space observations and for the use of arrays. In many applications the readout can increase sensitivity. The authors present noise spectra for 4He temperature bolometers with no excess noise at frequencies greater than 0.1 Hz. The measured optical responsivity of a bolometer operated with the present readout is the same as that of a bolometer operated with a conventional readout.
Document ID
19890048379
Acquisition Source
Legacy CDMS
Document Type
Reprint (Version printed in journal)
External Source(s)
Authors
Rieke, F. M.
(California Univ. Berkeley, CA, United States)
Lange, A. E.
(California Univ. Berkeley, CA, United States)
Beeman, J. W.
(California Univ. Berkeley, CA, United States)
Haller, E. E.
(California, University Berkeley, United States)
Date Acquired
August 14, 2013
Publication Date
February 1, 1989
Publication Information
Publication: IEEE Transactions on Nuclear Science
Volume: 36
ISSN: 0018-9499
Subject Category
Instrumentation And Photography
Accession Number
89A35750
Funding Number(s)
CONTRACT_GRANT: NSG-7205
CONTRACT_GRANT: DE-AC03-76SF-00098