NASA Logo

NTRS

NTRS - NASA Technical Reports Server

Press Enter or click the Search button to begin your search.

Back to Results
Laser wavemeter/spectrometer operating manualThe Laser Wavemeter/Spectrometer is a compact, rugged instrument using a unique combination of an uncoated (Snyder) wedge interferometer and a coated (Fizeau) wedge interferometer to simultaneously measure the wavelength centroid and spectral characteristics of pulsed or CW lasers. Wavelength centroid measurement accuracies are in the range of 1 part in 1,000,000 using the Snyder stage alone. The Fizeau wedge section is capable of resolving spectral details to 0.002/cm and of locating the position of a fringe to 0.0005/cm. Because the Snyder wedge is uncoated, wavelength centroid measurements are possible over the entire detector range of from about 300 nm to greater than 1.0 micron. The coatings of the Fizeau wedge limit spectral analysis to a 200 nm interval in that range. The instrument is currently configured to operate from 600 nm to 800 nm.
Document ID
19890001824
Acquisition Source
Legacy CDMS
Document Type
Contractor Report (CR)
Authors
Cotnoir, Leo J.
(Maryland Univ. College Park, MD, United States)
Date Acquired
September 5, 2013
Publication Date
February 1, 1986
Subject Category
Instrumentation And Photography
Report/Patent Number
NAS 1.26:180802
NASA-CR-180802
Report Number: NAS 1.26:180802
Report Number: NASA-CR-180802
Accession Number
89N11195
Funding Number(s)
CONTRACT_GRANT: NAS5-26654
Distribution Limits
Public
Copyright
Work of the US Gov. Public Use Permitted.
No Preview Available