Calibration technique for electromagnetic flowmetersThermal calorimetric method is used to calibrate electromagnetic flowmeters for liquid alkali metals. The electromagnetic flowmeter is placed in the liquid metal flow system in series with a thermal calorimeter. Therefore, the calculated flow rate through the calorimeter can be compared directly with the respective electromagnetic flowmeter reading.