NASA Logo

NTRS

NTRS - NASA Technical Reports Server

Press Enter or click the Search button to begin your search.

Back to Results
Active cavity radiometer, type III - An automatic, absolute standard, highly accurate detectorInstrument of simple construction operates without vacuum enclosure over wide pressure range and temperatures from 218 to 398 deg K and defines absolute radiometric scale to within less than 0.5 mW/sq cm. It has potential application to meteorology and climatology and operates on electrical substitution calorimeter principle.
Document ID
19710000130
Acquisition Source
Legacy CDMS
Document Type
Other - NASA Tech Brief
Authors
Willson, R. C.
Date Acquired
August 6, 2013
Publication Date
May 1, 1971
Subject Category
Physical Sciences
Report/Patent Number
NPO-11504
Report Number: NPO-11504
Accession Number
71B10131
Distribution Limits
Public
Copyright
Work of the US Gov. Public Use Permitted.
No Preview Available