NASA Logo

NTRS

NTRS - NASA Technical Reports Server

Press Enter or click the Search button to begin your search.

Back to Results
Development of techniques and associated instrumentation for high temperature emissivity measurementsThe calibrating and testing procedures for the radiometric emittance measurement apparatus are described. Test results are given for the NBS platinum-rhodium standard (SRM No. 1409), the NBS oxidized Kanthal standard, and the oxidized Rene 41 materials. Results indicate that with the platinum-rhodium standard the measurement system compared with the NBS data to within 0.02 for both total and spectral normal emitttance. Data from the Kanthal strip was consistently higher than NBS values. For oxidized Rene 41, no significant changes in as prepared emittance was observed. The emittance was stable for temperatures up to 1150 K.
Document ID
19730004694
Acquisition Source
Legacy CDMS
Document Type
Contractor Report (CR)
Authors
Cunnington, G. R.
(Lockheed Missiles and Space Co. Palo Alto, CA, United States)
Funai, A. I.
(Lockheed Missiles and Space Co. Palo Alto, CA, United States)
Date Acquired
September 2, 2013
Publication Date
September 28, 1972
Subject Category
Instrumentation And Photography
Report/Patent Number
LMSC-D309948
NASA-CR-123968
Report Number: LMSC-D309948
Report Number: NASA-CR-123968
Accession Number
73N13421
Funding Number(s)
CONTRACT_GRANT: NAS8-26304
Distribution Limits
Public
Copyright
Work of the US Gov. Public Use Permitted.