Cardiac R-wave detector with automatic sensitivity controlAn electronic circuit that automatically changes its sensitivity was developed for detecting the bioelectric signal resulting from activation of the heart's ventricles. Regulation of sensitivity was accomplished with two feedback channels that maintain the sensitivity level between an upper and a lower limit. These limits are proportional to the R-wave amplitude. Tests on an experimental circuit demonstrated a capability to reject unwanted signal noise, illustrated difficulties encountered without sensitivity control, and presented closed loop transients which occurred for step changes in R-wave amplitude.