Calibration of the electron-proton spectrometerThe principal function of the sensor used in the electron-proton spectrometer is to provide a signal which can be used to determine the energy and indicate the type of an incident particle. Two techniques are employed to resolve the particle intensity in different energy regions. The first employs a moderator surrounding each detector to provide a nominal lower limit to the energy of a particle which can be detected. The second technique utilizes a pulse height discriminator to identify those particles entering a detector whose energy is (1) sufficiently high that it exceeds the discriminator level if the particle is stopped in the detector, or (2) sufficiently low that the ionization rate causes the discrimination level to be exceeded for paths through the detector shorter than the particle range.