A narrowband, crystal controlled biomedical telemetry system.A miniature, single-channel, crystal-controlled transmitter has been developed for biomedical applications. A narrow-band frequency modulation (plus or minus 7 kHz) of the RF is used to achieve maximum operating range with minimum transmitting power. The radiated power is limited to stay within the 50 microvolts/m at 15 m FCC requirement for low-power transmitters in the 88- to 108-MHz band. This technique offers a number of advantages. First only manufacturer's type approval is required for the device and the user does not need a license. Second maximum operating range can be achieved within the FCC power limitations by using a narrow-band that is consistent with the required medical information bandwidth. A third advantage in using the 88- to 108-MHz band is that the commercial FM stations are relatively widely spaced and carefully regulated so that by selecting the transmitter's frequency in the guard band between stations, minimum interference from other transmitter sources is encountered.
Document ID
19730038579
Acquisition Source
Legacy CDMS
Document Type
Conference Proceedings
Authors
Westbrook, R. M.
Fryer, T. B. (NASA Ames Research Center Moffett Field, Calif., United States)