Three stage gas dilution system for calibration of air quality instrumentationTests were conducted on a three stage gas dilution system to evaluate the performance of the system for preparation of gas mixtures in the range of concentrations from 1 to 200 ppm. The test gas used in the evaluation was carbon monoxide. Mixtures of carbon monoxide and air were prepared in the gas dilution system and analyzed on stream with a commercially available infrared CO detector. Results showed the gas dilution system to be accurate to within 5, 3, and 2 percent in the respective ranges of 1-20 ppm, 20-50 ppm, and 50-200 ppm.