Flight contaminant trace analyser. Phase 1: Chromatographic input systemThe development of a chromatographic column capable of resolving compounds associated with spacecraft atmospheres is presented. Consideration is given to sampling techniques, column parameters and operation, and column interface with a mass spectrometer. A capillary column coated with a mixture of polyalkylene glycols is found to provide the best selectivity for resolving multicomponent mixtures found in spacecraft atmospheres. Temperature programming and isothermal operation of the column are evaluated and it is found that temperature programming has a shorter analysis time for a given carrier gas flow rate and overall superior resolution. It is observed that hydrogen provides a 15% savings in analysis time over helium. Following the optimization of column operational parameters, a mixed phase Ucon capillary is prepared for evaluation during the column test period in which the test sample is automatically analyzed. Analysis of the multicomponent test mixture is completed within 45 minutes provided temperature programming is used. All but two of the test compounds are well resolved.