NASA Logo

NTRS

NTRS - NASA Technical Reports Server

Press Enter or click the Search button to begin your search.

Back to Results
Flight contaminant trace analyser. Phase 1: Chromatographic input systemThe purpose of this investigation was to develop two chromatographic columns which would enable a mass spectral identification of 40 specified compounds. The columns are for use in a toxic gas analyzer, which incorporates an automated gas chromatograph-mass spectrometer. Different types of stationary phases were investigated. The columns used were of the open tubular capillary type and were made of nickel. Limitations of initial and final temperature of operation led to final development of a column which could resolve most of the compounds required. The few unresolved components are capable of resolution and identification by the mass spectrometer. The columns (182m Ni x 0.8m 0.D x 0.5mm I.D) coated with Witconal La 23, yielded in excess of 200,000 theoretical plates and completed the analysis in less than 90 minutes using a carrier gas flow rate of 4 cc/min hydrogen.
Document ID
19760024218
Acquisition Source
Legacy CDMS
Document Type
Contractor Report (CR)
Authors
Zlatkis, A.
(Ionics Research, Inc. Houston, TX, United States)
Date Acquired
September 3, 2013
Publication Date
September 1, 1976
Subject Category
Inorganic And Physical Chemistry
Report/Patent Number
NASA-CR-150942
Report Number: NASA-CR-150942
Accession Number
76N31306
Funding Number(s)
CONTRACT_GRANT: NAS9-14763
Distribution Limits
Public
Copyright
Work of the US Gov. Public Use Permitted.
No Preview Available