An investigation of accelerating mode and decelerating mode constant-momentum mass spectrometry and their application to a residual gas analyzerA theoretical analysis of constant momentum mass spectrometry was made. A maximum resolving power for the decelerating mode constant momentum mass spectrometer was shown theoretically to exist for a beam of ions of known energy. A vacuum system and an electron beam ionization source was constructed. Supporting electronics for a residual gas analyzer were built. Experimental investigations of various types of accelerating and decelerating impulsive modes of a constant momentum mass spectrometer as applied to a residual gas analyzer were made. The data indicate that the resolving power for the decelerating mode is comparable to that of the accelerating mode.
Document ID
19770009187
Acquisition Source
Legacy CDMS
Document Type
Contractor Report (CR)
Authors
Ng, Y. S.
(Pennsylvania State Univ. University Park, PA, United States)
Date Acquired
September 3, 2013
Publication Date
January 20, 1977
Subject Category
Inorganic And Physical Chemistry
Report/Patent Number
PSU-IRL-SCI-452
NASA-CR-149466
Report Number: PSU-IRL-SCI-452
Report Number: NASA-CR-149466
Accession Number
77N16130
Funding Number(s)
CONTRACT_GRANT: NGR-39-009-232
Distribution Limits
Public
Copyright
Work of the US Gov. Public Use Permitted.