Self-scanned anode array with a microchannel plate electron multiplier - The SSANACONA photoelectron counting detector utilizing a 128-element linear self-scanned anode array to collect the output of a dual microchannel plate electron multiplier is described. The operation of the detector, its spatial resolution and detection efficiency, and radiation behavior are discussed. The low-noise detector was developed for the Mariner Jupiter/Saturn mission UV spectrometer, and offers a large dynamic range and count rate of 0.003 per anode-sec. It can measure simultaneously two spatially separated signals differing by more than six orders of magnitude in count rate.