Tissue stimulator enclosure welding fixtureIt was demonstrated that the thickness of the stimulator titanium enclosure is directly related to the battery recharge time cycle. Reduction of the titanium enclosure thickness from approximately 0.37 mm (0.015 inch) to 0.05 mm (0.002 inch) significantly reduced the recharge time cycle and thereby patient inconvenience. However, fabrication of titanium enclosures from the thinner material introduced problems in forming, holding, and welding that required improvement in state of the art shop practices. The procedures that were utilized to resolve these fabrication problems are described.
Document ID
19780016612
Acquisition Source
Legacy CDMS
Document Type
Technical Memorandum (TM)
Authors
Mcclure, S. R.
(NASA Goddard Space Flight Center Greenbelt, MD, United States)
Date Acquired
September 3, 2013
Publication Date
October 1, 1977
Subject Category
Mechanical Engineering
Report/Patent Number
NASA-TM-79529
Report Number: NASA-TM-79529
Accession Number
78N24555
Distribution Limits
Public
Copyright
Work of the US Gov. Public Use Permitted.