X-ray polarimetersThe paper presents the theory, design, performance parameters, and statistical limitations of dispersive-crystal and Thomson-scattering devices as X-ray polarimeters for solar and extrasolar observations. With reference to the Bragg crystal polarimeter, the instrument parameters of the OSO-8 polarimeter are reported along with estimates of the minimum detectable polarization for several sources as derived from measured values for the integral reflectivities, the known spectra for a number of X-ray sources, and crystal panel geometry. For the Thomson-scattering polarimeter, values are given for the scattering efficiency for hydrogen and lithium as a function of photon energy and for the spectra of the two brightest X-ray sources: Sco X-1 and the Crab Nebula.
Document ID
19790027941
Acquisition Source
Legacy CDMS
Document Type
Conference Proceedings
Authors
Silver, E. H. (Columbia Univ. New York, NY, United States)
Novick, R. (Columbia Univ. New York, NY, United States)
Weisskopf, M. C. (Columbia University New York, N.Y., United States)
Date Acquired
August 9, 2013
Publication Date
January 1, 1977
Subject Category
Instrumentation And Photography
Meeting Information
Meeting: Optical polarimetry: Instrumentation and applications; Seminar