High-speed computerized tomographyThe development of a high-speed reconstruction processor and a channelized architecture to use with a high-resolution tomographic unit is discussed with attention to the convolution reconstruction algorithm. By means of this algorithm, input data and intermediate result precision required throughout the algorithm execution have been studied with computer simulation using profile data derived from mathematically simulated test objects and experimental animal data. A prototype section for a highly parallel all-digital system executes 60 million arithmetic operations per second, and the full-scale version is expected to reconstruct 500 to 1000 cross sections per second.
Document ID
19790028017
Acquisition Source
Legacy CDMS
Document Type
Conference Proceedings
Authors
Swartzlander, E. E., Jr. (TRW Defense and Space Systems Group Redondo Beach, Calif., United States)
Gilbert, B. K. (Mayo Foundation Rochester, Minn., United States)
Date Acquired
August 9, 2013
Publication Date
January 1, 1977
Subject Category
Man/System Technology And Life Support
Meeting Information
Meeting: Applications of digital image processing; International Optical Computing Conference, San Diego