NASA Logo

NTRS

NTRS - NASA Technical Reports Server

Press Enter or click the Search button to begin your search.

Back to Results
Modified displacement gage for cryogenic testingModification of double-cantilever-beam resistance strain gage makes boiling of hydrogen on gage arms less of problem. Modified gages are encapsulated nickel/chromium alloy, and bridge-excitation voltage is reduced from 10 to 1.5 volts. Sensitivity is 1.0 millivolt per inch with 1.5 volt excitation.
Document ID
19800000077
Acquisition Source
Legacy CDMS
Document Type
Other - NASA Tech Brief
Authors
Pierce, W. S.
Date Acquired
August 10, 2013
Publication Date
August 1, 1980
Publication Information
Publication: NASA Tech Briefs
Volume: 5
Issue: 1
ISSN: 0145-319X
Subject Category
Mechanics
Report/Patent Number
LEW-13039
Report Number: LEW-13039
ISSN: 0145-319X
Accession Number
80B10077
Distribution Limits
Public
Copyright
Work of the US Gov. Public Use Permitted.

Available Downloads

There are no available downloads for this record.