NASA Logo

NTRS

NTRS - NASA Technical Reports Server

Back to Results
Testing Evaluation of the Electrochemical Organic Content AnalyzerThe breadboard electrochemical organic content analyzer was evalauted for aerospace applications. An awareness of the disadvantages of expendables in some systems resulted in an effort to investigate ways of reducing the consumption of the analyzer's electrolyte from the rate of 5.17 kg/30 days. It was found that the electrochemical organic content analyzer can result in an organic monitor in the water quality monitor having a range of 0.1 to 100 mg/1 total organic carbon for a large number of common organic solutes. In a flight version it is anticipated the analyzer would occupy .0002 cu m, weigh 1.4 kg, and require 10 W or less of power. With the optimum method of injecting electrolyte into the sample (saturation of the sample with a salt) it would expend only 0.04 kg of electrolyte during 30 days of continuous operation.
Document ID
19800013495
Acquisition Source
Legacy CDMS
Document Type
Contractor Report (CR)
Authors
Davenport, R. J.
(Life Systems, Inc. Cleveland, OH, United States)
Date Acquired
September 4, 2013
Publication Date
November 1, 1979
Subject Category
Man/System Technology And Life Support
Report/Patent Number
LSI-TR-334-4
NASA-CR-160569
Report Number: LSI-TR-334-4
Report Number: NASA-CR-160569
Accession Number
80N21981
Funding Number(s)
CONTRACT_GRANT: NAS9-15402
Distribution Limits
Public
Copyright
Work of the US Gov. Public Use Permitted.
No Preview Available