ECC ozonesonde performance at high altitudes: Pump efficiencyThe ECC ozonesonde sampling behavior was examined at pressures ranging from 60 to 6 hPa with the objective of evaluating uncertainties in high altitude ozone data caused by variations in pumping efficiency. The averaged pump efficiency correction curve for a 43 sample set of 3A type ECC pumps showed a 2-3% bias from the curve provided by the manufacturer. In addition, random pump to pump variations (2 sigma) + or - were 5% at 6 hPa. These values probably represent minimum errors since the pumps were from the same production batch. A seven sample set of the newer 4A type ECC pumps was examined, with similar findings as for the 3A types.