The evaluation of a deformable diffraction grating for a stigmatic EUV spectroheliometerTwo crucial areas in the development of the high resolution EUV spectroheliometer are the elastically deformable toroidal diffraction grating and the open-structure imaging pulse counting detector system. The optical performance of a high frequency toroidal grating at EUV wavelengths was verified with a 3600 groove/mm master grating on a spherical concave blank. An open-structure (256 x 1024) pixel, multi-anode microchannel array was developed and tested as a pulse counting detector system. These two systems verify the proof of concept of the spectroheliometer design.