A position sensitive phoswich hard X-ray detector systemA prototype position sensitive phoswich hard X-ray detector, designed for eventual astronomical usage, was tested in the laboratory. The scintillation crystal geometry was designed on the basis of a Monte Carlo simulation of the internal optics and includes a 3mm thick NaI(T1) primary X-ray detector which is actively shielded by a 20 mm thick CsI(T1) scintillation crystal. This phoswich arrangement is viewed by a number two inch photomultipliers. Measured values of the positional and spectral resolution of incident X-ray photons are compared with calculation.