Organics in water contamination analyzer, phase 1The requirements which would result in identifying the components of an automatic analytical system for the analysis of specific organic compounds in the space station potable water supply are defined. The gas chromatographic system for such an analysis is limited to commercially available off-the-shelf hardware and includes the sample inlet, an ionization detector, capillary columns as well as computerized compound identification. The sampling system will be a special variation of the purge and trap Tenax mode using six-port valves and a 500 microliter water sample. Capillary columns used for the separating of contaminants will be bonded phase fused silica with a silicone stationary phase. Two detectors can be used: photoionization and far ultraviolet, since they are sensitive and compatible with capillary columns. A computer system evaluation and program with the principle of compound identification based on the retention index is presented.
Document ID
19870002627
Acquisition Source
Legacy CDMS
Document Type
Contractor Report (CR)
Date Acquired
September 5, 2013
Publication Date
June 2, 1986
Subject Category
Environment Pollution
Report/Patent Number
NASA-CR-179901
NAS 1.26:179901
Report Number: NASA-CR-179901
Report Number: NAS 1.26:179901
Accession Number
87N12060
Funding Number(s)
CONTRACT_GRANT: NAS9-17536
Distribution Limits
Public
Copyright
Work of the US Gov. Public Use Permitted.