NASA Logo

NTRS

NTRS - NASA Technical Reports Server

Back to Results
Low pressure gas flow analysis through an effusive inlet using mass spectrometryA mass spectrometric method for analyzing flow past and through an effusive inlet designed for use on the tethered satellite and other entering vehicles is discussed. Source stream concentrations of species in a gaseous mixture are determined using a calibration of measured mass spectral intensities versus source stream pressure for standard gas mixtures and pure gases. Concentrations are shown to be accurate within experimental error. Theoretical explanations for observed mass discrimination effects as they relate to the various flow situations in the effusive inlet and the experimental apparatus are discussed.
Document ID
19880012029
Acquisition Source
Legacy CDMS
Document Type
Contractor Report (CR)
Authors
Brown, David R.
(Old Dominion Univ. Norfolk, VA, United States)
Brown, Kenneth G.
(Old Dominion Univ. Norfolk, VA, United States)
Date Acquired
September 5, 2013
Publication Date
May 1, 1988
Subject Category
Fluid Mechanics And Heat Transfer
Report/Patent Number
NAS 1.26:182712
NASA-CR-182712
Report Number: NAS 1.26:182712
Report Number: NASA-CR-182712
Accession Number
88N21413
Funding Number(s)
CONTRACT_GRANT: NAS1-17993
Distribution Limits
Public
Copyright
Work of the US Gov. Public Use Permitted.
No Preview Available