Automated data acquisition and processing for a Hohlraum reflectometerA computer and data acquisition board were used to automate a Perkin-Elmer Model 13 spectrophotometer with a Hohlraum reflectivity attachment. Additional electronic circuitry was necessary for amplification, filtering, and debouncing. The computer was programmed to calculate spectral emittance from 1.7 to 14.7 micrometers and also total emittance versus temperature. Automation of the Hohlraum reflectometer reduced the time required to determine total emittance versus temperature from about three hours to about 40 minutes.
Document ID
19890005045
Acquisition Source
Legacy CDMS
Document Type
Technical Memorandum (TM)
Authors
Difilippo, Frank
(Case Western Reserve Univ. Cleveland, OH., United States)
Mirtich, Michael J.
(NASA Lewis Research Center Cleveland, OH, United States)
Date Acquired
September 5, 2013
Publication Date
December 1, 1988