Integrating sphere and hemisphere calibration updateThis background information relates to the conclusions reached about the GSFC/Optronic calibrations for the 6-foot sphere. Both the hemispherical and the spherical integrating sources were remeasured in February 1988 by two experimenters and the results show a decrease in the output of the sphere. But the discrepancy between GSFC data and Optronic Data remains. The decrease in the output can be accounted for by a correction in the current value of power supplies Number 2 and Number 3. The supplies were running about 1.75 percent high. Minor differences were noted in the short wavelength end of the hemisphere. Efforts to resolve the discrepancy between GSFC and Optronic Labs as well as improve the ability to make more accurate and precise measurements are underway.