NASA Logo

NTRS

NTRS - NASA Technical Reports Server

Press Enter or click the Search button to begin your search.

Back to Results
AVIRIS scan drive design and performanceThe Airborne Visible/Infrared Imaging Spectrometer (AVIRIS) images the ground with an instantaneous field of view (IFOV) of 1 mrad. The IFOV is scanned 30 deg from left to right to provide the cross-track dimension of the image, while the aircraft's motion provides the along-track dimension. The scanning frequency is 12 Hz, with a scan efficiency of 70 percent. The scan mirror has an effective diameter of 5.7 in, and its positional accuracy is a small fraction of a milliradian of the nominal position-time profile. Described are the design and performance of the scan drive mechanism. Tradeoffs among various approaches are discussed, and the reasons given for the selection of the cam drive.
Document ID
19890022948
Acquisition Source
Legacy CDMS
Document Type
Conference Paper
Authors
Miller, D. C.
(California Institute of Technology Jet Propulsion Laboratory, Pasadena, United States)
Date Acquired
August 13, 2013
Publication Date
January 1, 1988
Subject Category
Instrumentation And Photography
Meeting Information
Meeting: Imaging Spectroscopy II
Location: San Diego, CA
Country: United States
Start Date: August 20, 1987
End Date: August 21, 1987
Accession Number
89A10319
Distribution Limits
Public
Copyright
Other

Available Downloads

There are no available downloads for this record.