Contamination control of the Cryogenic Limb Array Etalon SpectrometerThe cleanliness requirements of the Cryogenic Limb Array Etalon Spectrometer (CLAES) are described and the control program is assessed with particular emphasis given to the effects of particulate interference. The use of the CLAES is reviewed, and off-axis scatter from the earth is identified as an important noise source. Since the CLAES altitude of interest causes a low-altitude detection of 0.2 deg above the earth's surface, straylight controls are important in the development of the instrument. The development of off-axis rejection and the particulate-cleanliness level are evaluated in terms of those required for a certain BRDF. The control plan is described, and procedures are outlined including cleaning methods and facilities. The modeling predictions and procedural controls are found to effectively satisfy requirements for low scatter levels and out-of-field view-rejection by maximizing straylight control and allowing a margin against noise-equivalent radiance.
Document ID
19910070357
Acquisition Source
Legacy CDMS
Document Type
Conference Paper
Authors
Steakley, B. C. (Lockheed Missiles and Space Co. Palo Alto, CA, United States)
King, A. D. (Lockheed Missiles and Space Co. Palo Alto, CA, United States)
Rigney, T. E. (Lockheed Missiles and Space Co., Inc. Research and Development Div., Palo Alto, CA, United States)
Date Acquired
August 14, 2013
Publication Date
January 1, 1990
Subject Category
Spacecraft Design, Testing And Performance
Meeting Information
Meeting: Optical System Contamination: Effects, Measurement, Control II