NASA Logo

NTRS

NTRS - NASA Technical Reports Server

Press Enter or click the Search button to begin your search.

Back to Results
Temperature Stabilization Requirements for Unchopped Thermal DetectorsThe temperature stabilization requirements of unchopped thermistor bolometers and thermopile detectors are analyzed. The detector temperature, on which the bolometer output signal depends, is quite sensitive to changes in instrument temperature but relatively insensitive to changes in scene temperature. In contrast, the difference in temperature between detector and substrate (instrument), on which the thermopile signal depends, is equally sensitive to changes in instrument and scene temperature. Expressions for these dependencies are derived based on a simplified instrument model. It is shown that for a typical uncooled thermal imager, the temperature stabilization requirements for a bolometer are two orders of magnitude more stringent than those for a thermopile detector. Keywords: thermistor, bolometer, thermopile, detector, IR, thermal, temperature stabilization
Document ID
20000050237
Acquisition Source
Headquarters
Document Type
Other
Authors
Foote, Marc C.
(Jet Propulsion Lab., California Inst. of Tech. Pasadena, CA United States)
Date Acquired
September 7, 2013
Publication Date
January 1, 2000
Subject Category
Instrumentation And Photography
Distribution Limits
Public
Copyright
Work of the US Gov. Public Use Permitted.
No Preview Available