NASA Logo

NTRS

NTRS - NASA Technical Reports Server

Press Enter or click the Search button to begin your search.

Back to Results
MECA Electrometer: An Old Instrument Finds a New ApplicationThe Mars '01 binder contains an electrometer designed to evaluate the electrostatic nature of the Martian regolith (soil) and atmosphere. The electrometer is part of MECA (Mars Environmental Compatibility Assessment) project. The objective is to gain a better understanding of the hazards related to the human exploration of Mars. The electrometer was designed to fit into the heel of the scoop of the robotic arm. The instrument has four sensor types: (a) triboelectric field, (b) electric-field, (c) ion current, and (d) temperature. The triboelectric field sensor array contains five insulating materials to determine material charging effects as the scoop is dragged through the Martian regolith. The sensor has an electric field sensitivity of 35 kV/cm-V and room temperature drift of approx.3 micro V/sec. The sensor has been operated as low as -60 C where the drift is undetected. This talk with focus on the operating principles of the triboelectric sensors and present some of test results.
Document ID
20000075639
Acquisition Source
Jet Propulsion Laboratory
Document Type
Other
Authors
Buehler, Martin
(Jet Propulsion Lab., California Inst. of Tech. Pasadena, CA United States)
Date Acquired
August 19, 2013
Publication Date
January 1, 2000
Subject Category
Lunar And Planetary Science And Exploration
Distribution Limits
Public
Copyright
Work of the US Gov. Public Use Permitted.

Available Downloads

There are no available downloads for this record.