NASA Logo

NTRS

NTRS - NASA Technical Reports Server

Press Enter or click the Search button to begin your search.

Back to Results
Cryogenic Pressure Calibrator for Wide Temperature Electronically Scanned (ESP) Pressure ModulesElectronically scanned pressure (ESP) modules have been developed that can operate in ambient and in cryogenic environments, particularly Langley's National Transonic Facility (NTF). Because they can operate directly in a cryogenic environment, their use eliminates many of the operational problems associated with using conventional modules at low temperatures. To ensure the accuracy of these new instruments, calibration was conducted in a laboratory simulating the environmental conditions of NTF. This paper discusses the calibration process by means of the simulation laboratory, the system inputs and outputs and the analysis of the calibration data. Calibration results of module M4, a wide temperature ESP module with 16 ports and a pressure range of +/- 4 psid are given.
Document ID
20010108002
Acquisition Source
Langley Research Center
Document Type
Technical Memorandum (TM)
Authors
Faulcon, Nettie D.
(NASA Langley Research Center Hampton, VA United States)
Date Acquired
September 7, 2013
Publication Date
October 1, 2001
Subject Category
Aerodynamics
Report/Patent Number
NAS 1.15:211031
NASA/TM-2001-211031
L-18098
Report Number: NAS 1.15:211031
Report Number: NASA/TM-2001-211031
Report Number: L-18098
Funding Number(s)
PROJECT: RTOP 922-20-80-01
Distribution Limits
Public
Copyright
Work of the US Gov. Public Use Permitted.