Material Outgassing, Identification and Deposition, MOLIDEP SystemThe outgassing tests are performed employing a modified vacuum operated Cahn analytical microbalance, identified as the MOLIDEP system. The test measures under high vacuum, the time varying Molecular mass loss of a material sample held at a chosen temperature; it Identifies the outgassing molecular components using an inline SRS 300 amu Residual Gas Analyzer (RGA) and employs a temperature controlled 10 MHz Quartz Crystal Microbalance (QCM) to measure the condensable DEPosits. Both the QCM and the RGA intercept within the conductive passage the outgassing products being evacuated by a turbomolecular pump. The continuous measurements of the mass loss, the rate of loss, the sample temperature, the rate of temperature change, the QCM temperature and the QCM recorded condensable deposits or rate of deposits are saved to an Excel spreadsheet. A separate computer controls the RGA.