Laboratory Astrophysics Using a Microcalorimeter and Bragg Crystal Spectrometer on an Electron Beam Ion TrapWe completed modifications to the new microcalorimeter system dedicated for use on the EBIT at NIST, which included: 1) a redesign of the x-ray calibration source from a direct electron impact source to one that irradiates the microcalorimeter with fluorescent x-rays. The resulting calibration lines are free of bremsstrahlung background; 2) the microcalorimeter electronic circuit was significantly improved to ensure long-term stability for lengthy experimental runs
Document ID
20050176542
Acquisition Source
Goddard Space Flight Center
Document Type
Contractor or Grantee Report
Authors
Brinton, John
(NASA Wallops Flight Center Wallops Island, VA, United States)
Silver, Eric
(Smithsonian Astrophysical Observatory Cambridge, MA, United States)
Date Acquired
September 7, 2013
Publication Date
May 1, 2005
Subject Category
Astrophysics
Funding Number(s)
CONTRACT_GRANT: NAG5-5362
Distribution Limits
Public
Copyright
Work of the US Gov. Public Use Permitted.