NASA Logo

NTRS

NTRS - NASA Technical Reports Server

Press Enter or click the Search button to begin your search.

Back to Results
Study of Static Microchannel Plate Saturation Effects for the Fast Plasma Investigation Dual Electron Spectrometers on NASA's Magnetospheric MultiScale MissionImaging detecting systems based on microchannel plates (MCPs) are the most common for low energy plasma measurements for both space borne and ground applications. One of the key parameters of these detection systems is the dynamic range of the MCP's response to the input fluxes of charged particles. For most applications the dynamic range of the linear response should be as wide as possible. This is especially true for the Dual Electron Spectrometers (DESs) of the Fast Plasma Investigation (FPI) on NASA's Magnetospheric MultiScale (MMS) mission because a wide range of input fluxes are expected. To make use of the full available dynamic range, it is important to understand the MCP response behavior beyond the linear regime where the MCPs start to saturate. We have performed extensive studies of this during the characterization and calibration of the DES instruments and have identified several saturation effects of the detection system. The MCP itself exhibits saturation when the channels lack the ability to replenish charge sufficiently rapidly. It is found and will be shown that the ground system can significantly impact the correct measurement of this effect. As the MCP starts to saturate, the resulting pulse height distribution (PHD) changes shape and location (with less pulse height values), which leads to truncation of the PHD by the threshold set on the detection system discriminator. Finally, the detection system pulse amplifier exhibits saturation as the input flux drives pulse rates greater than its linear response speed. All of these effects effectively change the dead time of the overall detection system and as a result can affect the quality and interpretation of the flight data. We present results of detection system saturation effects and their interaction with special emphasis on the MCP related effects.
Document ID
20150011080
Acquisition Source
Goddard Space Flight Center
Document Type
Abstract
Authors
Avanov, L. A.
(Maryland Univ. College Park, MD, United States)
Gliese, U.
(SGT, Inc. Greenbelt, MD, United States)
Pollock, C. J.
(NASA Goddard Space Flight Center Greenbelt, MD, United States)
Moore, T. E.
(NASA Goddard Space Flight Center Greenbelt, MD, United States)
Chornay, D. J.
(Maryland Univ. College Park, MD, United States)
Barrie, A. C.
(SGT, Inc. Greenbelt, MD, United States)
Kujawski, J. T.
(NASA Goddard Space Flight Center Greenbelt, MD, United States)
Gershman, D. J.
(Oak Ridge Associated Universities Greenbelt, MD, United States)
Tucker, C. J.
(Global Science and Technology, Inc. Greenbelt, MD, United States)
Mariano, A.
(NASA Goddard Space Flight Center Greenbelt, MD, United States)
Smith, D. L.
(Orbital Sciences Corp. Greenbelt, MD, United States)
Jacques, A. D.
(NASA Goddard Space Flight Center Greenbelt, MD, United States)
Date Acquired
June 18, 2015
Publication Date
April 20, 2015
Subject Category
Instrumentation And Photography
Space Sciences (General)
Report/Patent Number
GSFC-E-DAA-TN21258
Report Number: GSFC-E-DAA-TN21258
Meeting Information
Meeting: Measurement Techniques in Solar and Space Physics
Location: Boulder, CO
Country: United States
Start Date: April 20, 2015
End Date: April 24, 2015
Sponsors: NASA Headquarters
Funding Number(s)
CONTRACT_GRANT: NNH06CC03B
CONTRACT_GRANT: NNG12CR31C
CONTRACT_GRANT: NNG06EO90A
CONTRACT_GRANT: NNG12CR29C
CONTRACT_GRANT: NNG11PL02A
Distribution Limits
Public
Copyright
Public Use Permitted.
Keywords
detecting system
microchannel plates
space plasma
No Preview Available