Final Report Document: Microgravity Medical Eyewash BSenior Capstone Design Team 15 has been commissioned by the National Aeronautics and Space Administration (NASA) to redesign the current microgravity eye wash station. The current design exhibits four primary limitations: single-use operation, excessive mass and volume, operational complexity requiring coordination with external systems, and dependence on frequent Earth resupply missions. Our updated eyewash design will be deployed on long-range space missions that could last multiple years. Based on these and other requirements from our customer, we have developed our updated eyewash design. Our compact, single-eye system uses capillary-driven flow and a free-floating
water ball. This design allows us to eliminate pumps and bulky tubing, decreasing the mass and volume of the overall system. The containment eyecup is held to the astronaut's face with a strap, and a silicone seal prevents leakage while improving user comfort. Contaminated water is contained within a disposal bag containing wicking material to pull it back out of the eyecup after washing the eye. This system is reusable, efficient, and easy to activate quickly. We evaluated our design through calculations, microfluidic testing, and user try-on testing to verify our requirements. Our work demonstrated that steady-state flow was capable with our capillary framework, achieving 1.31 L/min. The prototype construction confirmed mass and volume reductions, and try-on-testing evaluated the ability to put on the system quickly. Almost all major design requirements were achieved, and water loss could be validated by testing in microgravity.
Our design went through a number of iterations to reach the final solution presented in this report. These changes were driven by our testing and collaboration from each member of the team. Updated models optimized tube placement and geometry of the eyecup to better direct flow as it pools across the eye. Changes to our tubing bends improved capillary efficiency and increased the flow rate we were capable of achieving. And improved bracket placements adjusted the fit, comfort, and seal of the eyecup to the astronaut's face. Our capillary-based eyewash system is technically feasible and has been theoretically validated to operate effectively in microgravity conditions. Testing and analysis indicate that it can meet or exceed the required performance metrics, including flow rate and safety constraints. Overall, the design represents a viable alternative to current ISS-dependent eyewash systems for future space missions.
Document ID
20260004954
Acquisition Source
Kennedy Space Center
Document Type
Contractor or Grantee Report
Authors
Jeffrey Allen
(Michigan Technological University Houghton, United States)
Cody A Haney
(Michigan Technological University Houghton, United States)
Charley L Isham
(Michigan Technological University Houghton, United States)
Anish Krishnamurthy
(Michigan Technological University Houghton, United States)
Andrew J Van Lente
(Michigan Technological University Houghton, United States)
Bhisham N Sharma
(Michigan Technological University Houghton, United States)
Date Acquired
June 1, 2026
Publication Date
April 22, 2026
Subject Category
Aerospace Medicine
Funding Number(s)
WBS: 566548
Distribution Limits
Public
Copyright
Use by or on behalf of the US Gov. Permitted.
Technical Review
NASA Technical Management
No Preview Available