Pressure effect on the sensitivity of quartz Bourdon tube gauges.The sensitivity change for a commercial fused quartz Bourdon tube precision pressure gauge, due to a change in absolute pressure level, has been analytically computed and experimentally confirmed. The computed differential pressure error is 2.5% of full scale at a 100 atm absolute pressure level. The experimental method compared the fused quartz Bourdon tube gauge digital output to the results obtained from a nitrogen gas pressure system which had a high pressure, well-type mercury manometer as the differential pressure reference.