Heat conduction errors and time lag in cryogenic thermometer installationsInstallation practices are recommended that will increase rate of heat exchange between the thermometric sensing element and the cryogenic fluid and that will reduce the rate of undesired heat transfer to higher-temperature objects. Formulas and numerical data are given that help to estimate the magnitude of heat-conduction errors and of time lag in response.
Document ID
19740028299
Document Type
Conference Paper
Authors
Warshawsky, I. (NASA Lewis Research Center Cleveland, Ohio, United States)
Date Acquired
August 7, 2013
Publication Date
October 1, 1973
Subject Category
INSTRUMENTATION AND PHOTOGRAPHY
Meeting Information
National Conference and Exhibit(Houston, TX)
Distribution Limits
Public
Copyright
Other