NASA Logo

NTRS

NTRS - NASA Technical Reports Server

Back to Results
Modification of the Varian XL-100 NMR spectrometer for submilligram natural abundance C-13 analysesQuadrature detection modifications and a microcell arrangement are described for the Varian XL-100 NMR spectrometer which routinely improve sensitivity in C-13 Fourier transform sample-limited studies by a factor of 4 to 5. The power requirement of the RF pulse amplifier is relaxed by a factor of 4. Previously attainable resolution is not affected.
Document ID
19740061631
Acquisition Source
Legacy CDMS
Document Type
Reprint (Version printed in journal)
External Source(s)
Authors
Wilson, D. M.
Olsen, R. W.
Burlingame, A. L.
(California, University Berkeley, Calif., United States)
Date Acquired
August 7, 2013
Publication Date
September 1, 1974
Publication Information
Publication: Review of Scientific Instruments
Volume: 45
Subject Category
Instrumentation And Photography
Accession Number
74A44381
Funding Number(s)
CONTRACT_GRANT: NGL-05-003-003
Distribution Limits
Public
Copyright
Other

Available Downloads

There are no available downloads for this record.
No Preview Available