NASA Logo

NTRS

NTRS - NASA Technical Reports Server

Back to Results
Quadrant anode image sensorA position-sensitive event-counting electronic readout system for microchannel plates (MCPs) is described that offers the advantages of high spatial resolution and fast time resolution. The technique relies upon a four-quadrant electron-collecting anode located behind the output face of the microchannel plate, so that the electron cloud from each detected event is partly intercepted by each of the four quadrants. The relative amounts of charge collected by each quadrant depend on event position, permitting each event to be localized with two ratio circuits. A prototype quadrant anode system for ion, electron, and extreme ultraviolet imaging is described. The spatial resolution achieved, about 10 microns, allows individual MCP channels to be distinguished.
Document ID
19770028987
Acquisition Source
Legacy CDMS
Document Type
Reprint (Version printed in journal)
External Source(s)
Authors
Lampton, M.
(California Univ. Berkeley, CA, United States)
Malina, R. F.
(California, University Berkeley, Calif., United States)
Date Acquired
August 8, 2013
Publication Date
November 1, 1976
Publication Information
Publication: Review of Scientific Instruments
Volume: 47
Subject Category
Instrumentation And Photography
Accession Number
77A11839
Funding Number(s)
CONTRACT_GRANT: NGR-05-003-450
Distribution Limits
Public
Copyright
Other

Available Downloads

There are no available downloads for this record.
No Preview Available