Portable He-3 detector cryostat for the far infraredThe design of a portable He-3 cryostat for far infrared detection applications is described, with a cutaway drawing of the cryostat, including bolometer and cooled optics, provided. Consideration is given to the selection and testing of various bolometer materials, including Ge:Ga and Ge:InSb; the resistance-temperature coefficients of the materials examined are presented. The absolute flux calibration of the detector system using a new temperature-modulated cold source method is described.