Measurement of the polarization effects of an instrument using partially polarized lightAccuracy of a radiometer is adversely affected by scene polarization if the receiving optical system is sensitive to polarization. It is therefore necessary to specify and measure the sensitivity of the system to polarized light. The Mueller-Stokes matrix of an instrument may be determined experimentally and used to predict the effects of the instrument on any beam. The specification of a maximum polarization sensitivity stated in terms of the degree of polarization produced in an unpolarized beam can be experimentally verified even though an unpolarized beam is not available in the laboratory for direct measurement.
Document ID
19790043967
Acquisition Source
Legacy CDMS
Document Type
Reprint (Version printed in journal)
Authors
Howell, B. J.
(NASA Goddard Space Flight Center Earth Observation Systems Div., Greenbelt, Md., United States)
Date Acquired
August 9, 2013
Publication Date
March 15, 1979
Publication Information
Publication: Applied Optics
Volume: 18
Subject Category
Instrumentation And Photography
Accession Number
79A27980
Distribution Limits
Public
Copyright
Other